In our all our Stopped-Flow instruments, syringe velocity is
programmed and thus is independent of the solution viscosity. Very
asymmetric mixing such as pure glycerol against water can therefore be
achieved.
Mixer technology The mixer is the heart of a stopped-flow instrument. Although being
often referred as "T-mixer" a straight T-connection between 3
tubes is capable of mixing two solutions only in limited cases (aqueous
solutions, symmetrical 1:1 mixing, limited range of flow rate). The
mixer used in the Bio-Logic Stopped-Flow are all based on the "ball
mixer" technology which is a proven design for creating turbulence
in the most stringent conditions. Because of it the Bio-Logic
Stopped-Flow can operate in a wide range of fluidic conditions such as :
high dilution ratio (up to 1:100), high range of flow rate (from 1 to 20
mL/s), mixing of highly viscous solutions against water (viscosity up to
2000 centipoises or 2 pascal.second have been achieved). |